1 continuous dosing, Continuous dosing – Metrohm 916 Ti-Touch User Manual
Page 273

■■■■■■■■■■■■■■■■■■■■■■
27 Manual control
916 Ti-Touch
■■■■■■■■
257
[Dosing fixed vol.]
Dose a specified volume (see Chapter 27.3.2, page 258).
[Dosing]
Dose continuously (see Chapter 27.3.1, page 257).
27.3.1
Continuous dosing
Manual control: Dosing
▶ Dosing
With the function [Dosing], dosing is continuous for as long as you hold
the button [Dose] pressed down.
Proceed as follows:
1 Configure the dosing function
NOTE
■
The dosing and filling rates should be decreased for viscous liq-
uids.
■
The maximum dosing and filling rates depend on the cylinder
volume (see Chapter 31, page 406).
■
Enter the dosing rate.
■
Enter the filling rate.
2 Start dosing
■
Tap on [Dose].
The volume dosed is displayed. When the volume of one cylinder has
been dosed, the dosing cylinder will be refilled automatically.