2 dosing fixed volumes, Dosing fixed volumes – Metrohm 916 Ti-Touch User Manual
Page 274

27.3 Dosing
■■■■■■■■■■■■■■■■■■■■■■
258
■■■■■■■■
916 Ti-Touch
3 Fill the cylinder
■
Tap on [Fill].
The dosing cylinder is filled. The displayed volume value will be reset
to 0.0000 mL.
NOTE
If you exit the dialog with the fixed key [
], the dosing cylinder
will be filled automatically.
Parameter description
Dosing rate
Rate at which it is dosed.
Input range
0.01 - 166.00 mL/min
Selection
maximum | dynamic
Default value
maximum
maximum
The maximum dosing rate depends on the cylinder volume (see Chap-
ter 31, page 406).
dynamic
This setting can only be selected when the dosing cylinder is filled. The
dosing is being carried out faster and faster until the maximum dosing
rate is reached (starting with 1 mL/min and then doubling the dosing
rate every 1.5 s).
Filling rate
Rate at which the dosing cylinder is filled.
Input range
0.01 - 166.00 mL/min
Selection
maximum
Default value
maximum
maximum
The maximum filling rate depends on the cylinder volume (see Chapter
31, page 406).
27.3.2
Dosing fixed volumes
Manual control: Dosing
▶ Dosing fixed vol.