Metrohm 916 Ti-Touch User Manual
Page 319

■■■■■■■■■■■■■■■■■■■■■■
28 Parameters
916 Ti-Touch
■■■■■■■■
303
Measuring mode pH:
Input range
0.10 - 9.99
Default value
0.50
Measuring mode U, Ipol:
Input range
1 - 999 mV
Default value
30 mV
Measuring mode Upol:
Input range
0.1 - 99.9 µA
Default value
2.0 µA
EP recognition
This parameter allows you to filter out only the equivalence points that are
being sought.
Selection
first | greatest | last | ascending | descending
Default value
first
first
Only the first equivalence point will be recognized.
greatest
Only the equivalence point with the greatest ERC value, i.e. the steep-
est jump, will be recognized.
last
Only the last equivalence point will be recognized.
ascending
Only equivalence points with a positive slope of the titration curve will
be recognized.
descending
Only equivalence points with a negative slope of the titration curve will
be recognized.
Only one equivalence point will be recognized in a window. The number-
ing of the equivalence points (EP) is defined by the numbering of the win-
dows (e.g. EP2 in window 2), so that even if EPs are missing, the calcula-
tions will still be carried out with the correctly assigned EP volumes.