Min. slope
Minimum slope of the pH electrode.
for very good electrodes:
Input range
0.1 - 999.9 %
Default value
96.5 %
for good electrodes:
Input range
0.1 - 999.9 %
Default value
96.0 %
for usable electrodes:
Input range
0.1 - 999.9 %
Default value
95.0 %
Max. slope
Maximum slope of the pH electrode.
for very good electrodes:
Input range
0.1 - 999.9 %
Default value
101.0 %
for good electrodes:
Input range
0.1 - 999.9 %
Default value
102.0 %
for usable electrodes:
Input range
0.1 - 999.9 %
Default value
103.0 %
Response time
The voltage measured after three minutes in stirred solution serves as a
comparison value for determining the response time. The response time is
the time after which the measured voltage is within ±1 mV of this com-
parison value.
for very good electrodes:
Input range
0 - 999 s
Default value
45 s
for good electrodes:
Input range
0 - 999 s
Default value
50 s