28.1 Dynamic equivalence point titrations (DET)

■■■■■■■■■■■■■■■■■■■■■■

278

■■■■■■■■

916 Ti-Touch

Dosing rate

Rate to be dosed with until the start measured value or the start slope is
reached.

Input range

0.01 - 166.00 mL/min

Default value

5.00 mL/min

Selection

maximum

maximum
The maximum dosing rate depends on the cylinder volume (see Chap-
ter 31, page 406)
.

Initial measured value
The conditions for the entry of the first measured value in the measuring
point list are defined under [Initial meas. value].

Signal drift

Maximum permissible drift for the measured value acceptance, i.e. maxi-
mum change of the measured value per minute. For this value, a suitable
maximum waiting time is automatically calculated except you already have
modified this waiting time.

Measuring mode pH, U and Ipol:
Input range

0.1 - 999.0 mV/min

Selection

off

Default value

off

off
Measured value acceptance will take place after the maximum waiting
time has elapsed.

Measuring mode Upol:
Input range

0.01 - 99.90 µA/min

Selection

off

Default value

off

off
Measured value acceptance will take place after the maximum waiting
time has elapsed.

Min. waiting time

The measured value is not accepted until the minimum waiting time has
elapsed, even if the signal drift has already been reached. The minimum
waiting time is only important for drift-controlled measurements.

Input range

0 - 999999 s

Default value