1 inserting a command, Inserting a command – Metrohm 916 Ti-Touch User Manual
Page 191

■■■■■■■■■■■■■■■■■■■■■■
16 Methods
916 Ti-Touch
■■■■■■■■
175
[Insert command]
Insert a new method command. It is inserted before the selected com-
mand.
NOTE
Not all commands are available for subsequences. Commands which
cannot be inserted into subsequences are disabled.
[Delete command]
Delete the selected method command.
[Edit command]
Edit the selected method command.
16.4.1
Inserting a command
A method is comprised of individual commands. When you start a deter-
mination, the commands are executed one after the other.
The following constraints apply:
■
Maximum of 99 commands
■
Maximum of nine commands for titrations, measurements and calibra-
tions
Inserting a command
To insert a command in a method, proceed as follows:
1 Open the command overview
■
In the main dialog, tap on [Edit parameters].
■
Tap on the button [Insert command].