Metrohm 916 Ti-Touch User Manual

Page 374

Advertising
background image

28.9 Calculations

■■■■■■■■■■■■■■■■■■■■■■

358

■■■■■■■■

916 Ti-Touch

Upper limit

When this value is exceeded, the action defined in the following is trig-
gered.

Input range

0 - 9999999999

Default value

9999999999

Action

Selection of the action, which is carried out when the result limits are
exceeded.

Selection

Display message | Document message | Cancel
determination

Default value

Display message

Display message
A message is displayed. You can select whether you want to continue
with the determination or cancel the run.
Document message
In the determination data it will be documented, that the result limits
have been exceeded.
Cancel determination
The determination is stopped.

Dialog "Edit calculation / Result options"
In the dialog Edit calculation / Result options, settings for how to
process the calculated result are defined.

Variable for mean value

If the statistic calculations have been activated (see method options), the
mean value of the single results will be saved as variable SMN1 to SMN9.
As default value, always the first free variable is being displayed.

Selection

SMN1 … SMN9 | off

off
For the result not statistic calculations are carried out.

Save as titer

The result can be saved as titer for the selected titrant.

Advertising