67XXU USB/Serial Scale Service Manual

3.7 Communication

3.8 Analog Circuitry

l

24-bit A/D converter

l

Conversion Speed: 10Hz

l

Input range: -15mV to +15mV

l

Output code:1mV input will output about 100,000 raw counts

l

Hardware low pass filter and 2 programmable digital low pass filters

3.9 Capacity and Division

l

Programmable

l

Max display range: -199,999 to 199,999

l

Division number range for primary unit: 100-20,000

m

Division number range for second unit: 100-25,000 (division number will
be limited by REGUL setting)

l

Recommended Sensitivity: >1µV/ display division

3.10 Accuracy

≤0.01%

3.11 Calibration Method

l

Software calibration with long-term storage in EEPROM

l

Provides smooth curve fit through four points

l

Calibration can be done under kg or lb weight unit with 10% -100%FS
standard weight.

l

Weight fine adjustment (±

10%)

l

Geographic adjustment

Serial port1

Full-duplex RS-232

Optional Serial Port

USB (Virtual RS-232)

Baud Rate

Selectable: 1200-2400-4800-9600-19200-38400 bps

Data Output Format

8N1, 7O1, 7E1

Protocol

programmable