2 geographical adjustment, Geographical adjustment – Salter Brecknell 6700U Series (USB) User Manual

Page 57

Advertising
background image

67XXU USB/Serial Scale Service Manual

57

6.2 Geographical Adjustment

1.

When GEO is selected, press the [TARE] key to enter Geographical
Adjustment.

2.

When

CODE is displayed, press the [UNIT] key to select geographical position

code or input user local gravity value directly.

3.

When CODE is selected, choose the position code of the scale being used
(00-70) according to the elevation and latitude from

Table 6.1

by using the

[HOLD] and [UNIT] keys.

4.

Press [TARE] key to confirm.

CAUTION! Only an authorized manufacturer representative or certified
verification personnel may make these changes. Changing the Geographical
setting alters the calibration values!