67XXU USB/Serial Scale Service Manual

79

8.2 Audit Trail Parameters

Two separate incrementing, non-resetable audit trail parameters are used to indicate
changes to various parameters or calibration.

8.2.1 View Configuration Counter

1.

Press and hold the [HOLD] + [ON/OFF] key for 3 seconds.

CONFG is

displayed.

2.

Press the [TARE] key to view the configuration counter.

2a.

The display will briefly show the number of times the configuration has been
changed and will save up to 9999. This counter can't be modified or erased. It
counts from 0000 to 9999. When 9999 is reached, the counter starts over at
0000. The display will then show whether the configuration switch is on
(

cf.on) or (cf.off) off.

To Exit:

1.

Press the [ON/OFF] key twice.

Exit? is displayed.

2.

Press the [TARE] key to return to the working mode.

8.2.2 View Calibration Counter

1.

Press and hold the [HOLD] + [ON/OFF] key for 3 seconds.

CONFG is

displayed.

2.

Use the [PRINT] key to select the CAL menu.

3.

Press the [TARE] key to enter the calibration mode.

3a.

After entering this mode, the number of calibrations will be shown first. This
number will be increment one digit after every time a calibration has been
performed and saved. This counter can't be modified or erased. It counts from
0000 to 9999, when 9999 is reached, the counter starts over at 0000.

4.

After the counter number was displayed, it will show "

Ca.OFF" or "CAL-ON"

which depends on whether the sealed calibration switch is OFF or ON. If the
switch is OFF, the following steps can be done, but the result will not be
saved.

To Exit:

1.

Press the [ON/OFF] key twice.

Exit? is displayed.

2.

Press the [TARE] key to return to the working mode.