Salter Brecknell 6700U Series (USB) User Manual

Page 42

Advertising
background image

42

67XXU USB/Serial Scale Service Manual

COM1

(USB)

BUD.RT

1200

9600

Select COM1 baud rate

2400

4800

9600

19200

38400

BT.FMT

8N1

8N1

Select COM1 byte format

(1) 8N1=8 data bits, No parity check
bit, 1 stop bit

7O1

(2) 7O1=7 data bits, 1 Odd parity
check bit, 1 stop bit

7E1

(3) 7E1=7 data bits, 1 Even parity
check bit, 1 stop bit

OUT.MD

none

PRt.cd

Select COM1 output mode

(1) NONE=no communication

CONT

(2) CONT=continuously output

PRINT

(3) PRINT=output after [PRINT] key
is pressed

CMD

(4) CMD=output after a request
command is received

PRT.CD

(5) PRT.CD= output after [PRINT]
key is pressed or request command
received

STABL

(6) STABL=Sends output
automatically after scale is stable.
Note: use PRINT or CMD to output
data, the scale must be stable

ST.NLD

**(7) ST.NLD=output after scale has
returned to zero range
(OTHER>NLD.RG) and is stable.

NLD

**(8) NLD= Output after scale has
returned to zero range
(OTHER>NLD.RG) and does not
wait for the scale to be stable.

Pr.nld

(9) PR.NLD= Manual push button
print once after scale is stable.
Scale has to return to zero range
(OTHER>NLD.RG) before another
push button print command can be
executed.

**Note: If PRINT, STABL, ST.NLD, LNLD or CMD are used to output data, the scale must be stable.

USER

SubMenu1 SubMenu2

Option

Default

Parameter Description

Comment

Advertising