67XXU USB/Serial Scale Service Manual

69

7.6.3 Parameter 8213-SCP03

Set the USER-COM-LYOUT parameter to

SCP03.

Command: W (57h), request current reading

Command: Z (5ah)

If ZERO function cannot be activated, it will return to current scale status.

Table 7.5 Status Bits

Response

<STX>W

1

W

2

<dp>W

3

W

4

W

5

<CR>---normal data

Note: (1) The decimal point position is determined by CONFG-PRIM.D
(2) If current unit is "lb:oz", the format will be similar with following:
<STX>W

1

W

2

W

3

lbW

4

W

5

<dp>W

6

oz<CR>

Response

Zero function is activated (simulate ZERO key) and it returns to current scale status.
<STX>?[status]<CR>

Bit

Status

0

0 = not in motion
1 = in motion

1

0 = not over capacity
1 = over capacity

2

0 = not under zero (positive polarity)
1 = under zero (negative polarity)

3

0 = inside zero capture range
1 = outside zero capture range

4

0 = not center of zero
1 = center of zero

5

always 1

6

always 1

7

parity