6.3 Gravity Value

1.

When

GRAVT is selected, press the [HOLD] and [UNIT] keys to input the

gravity value of the position that scale is used (9.76183-9.99999).

2.

Press the [TARE] key to confirm.

6.4 Input or View Calibration

1.

When

INPUT is selected, press the [TARE] key to enter Input calibration

parameter values that were previously received or view current calibration
parameter values.

All parameters regarding calibration are divided to 12 pages and are displayed on LCD
by "nn:xxxx" format ("nn" is a decimal number of page, "xxxx" is an hexadecimal value
of parameter. (e.g. 02:85E2).

01-02 pages: zero code

03-04 pages: standard weight of CAL.P1

05-06 pages: codes of CAL.P1

07-08 pages: standard weight of CAL.P2

09-10 pages: codes of CAL.P2

11-12 pages: full capacity net code

2.

When no digits are blinking on the display, this means calibration parameters
value are being viewed. Use the [HOLD] key to view the next page.

3.

When parameter values are being viewed, press the [HOLD] key for more
than 3 seconds to modify.

4.

When first digit is blinking, this means the value is being modified. Press the
[HOLD] key to make next digit flash (if current flashing position is the last one
the next page value will be shown).

5.

Press the [HOLD] and [UNIT] keys to input the number.

6.

Press the [TARE] key to confirm.

7.

After the indicator gets all needed data, it will calculate and store all calibration
parameters into EEPROM or after finishing calibration, it will display

CAL.nd.

8.

The scale will re-start and go back to original mode.

NOTE: Only an authorized manufacture representative or certified verification
personnel may make these changes.