67XXU USB/Serial Scale Service Manual

7.6.5 Parameter EH-SCP (PS-60)

Set the USER-COM-LYOUT parameter to

eh-scp

This protocol is compatible with Toledo PS60 protocol. The baud rate and data format
is set by User menu.

Output Status Bit Meaning

Table 7.7 Status Bits

Summary of Command and Response

Bit

Status

0

0 = not in motion
1 = in motion

1

0 = not over capacity
1 = over capacity

2

0 = not under zero (positive polarity)
1 = under zero (negative polarity)

3

0 = inside zero capture range
1 = outside zero capture range

4

0 = not center of zero
1 = center of zero

5

always 1

6

always 1

7

parity

Command

Response

ASCII

HEX

W

57

Read scale weight:

<STX>W

1

W

2

<DP>W

3

W

4

W

5

<CR>---normal data

<STX>?[status]<CR>--if current weight is invalid

S

53

Read scale status:
<STX>?[status]<CR>

Z

5a

Simulate ZERO key:
<STX>?[status]<CR>

T

54

Simulate TARE key:
<STX>?[status]<CR>

X

58

Simulate OFF key:
power off the scale

others

Unknown commands:
<STX>?[status]<CR>