2 com port 1, 3 com port 2, 4 protocol – Salter Brecknell 6700U Series (USB) User Manual

Page 62: 5 transaction string, Com port 1 com port 2 protocol transaction string

Advertising
background image

62

67XXU USB/Serial Scale Service Manual

7.2 Com Port 1

COM1 is a USB port used as a virtual RS-232 port. Communication wires come from
the USB connector and TXD0, RXD0 and GND.

7.3 Com Port 2

COM2 is a RS-232 bi-directional port. Communication wires are connected to the RS-
232 connector using TXD1, RXD1 and GND. Refer to

page 19

for connection details.

7.4 Protocol

The baud rate and byte format is set by USER-COM-BUD.RT and USER-COM-
BT.FMT.

Responses to serial commands will be immediate or within one weight measure cycle
of the scale. One second should be adequate for use as a time-out value by remote
(controlling) device.

7.5 Transaction String

The length of each item in a transition string:

l

Reading data --- 6 bytes

l

Data polarity ----1 byte: "-" for negative, and followed the first digit; " " for
positive.

l

Decimal point ---1 byte: "."

l

Measure unit ----1-5 bytes:" lb"," kg","lb:oz","pcs","%". Units are always
lower case, left aligned

l

Current status-- 4 bytes

m

If the weight is over capacity, the scale will return eight "^" characters
(the field of polarity, decimal point, weight data is filled by "^").

m

If the weight is under capacity, it will return eight "_" characters (the field
of polarity, decimal point, and weight data is filled by "_").

m

If the zero point is error, it will return eight "-" characters (the field of
polarity, decimal point, and weight data is filled by "-").

Useless leading 0 before digits is suppressed. Reading weight is right aligned.

Table 7.1 Symbols Used

<LF>

Line Feed character (hex 0AH)

<CR>

Carriage Return character (hex 0DH)

<ETX>

End of Text character (hex 03H)

<SP>

Space (hex 20H)

H

1

H

2

H

3

H

4

Four current status bytes

<P>

Polarity character: "-" or " "