67XXU USB/Serial Scale Service Manual

81

9.2 Display Symbols

Symbol

Description

0----

Zero is over the setting range

0_____

Zero point is below the setting range

Ad----

Signal to ADC is over maximum range

Ad____

Signal to ADC is below minimum range

----

Weight is over upper limitation or display data is over limitation

____

Weight is below lower limitation

EEP.E1

CONFG or CAL parameters are not correctly set

EEP.E2

USER parameter is not correctly set

Lo.bAt

Battery voltage is lower than setting

CAP.- - -

Next displaying content is capacity

CAP.ER

Parameters about Capacity is not correct

CAL.Px

Calibration on point (x)

CAL.OFF

Calibration seal switch is on OFF position

CAL.ON

Calibration seal switch is on ON position

CAL.Er

Calibration error, maybe input data or loaded weight is incorrect, unstable, non-linear

CA.End

End calibration

OFF

Power OFF the indicator

STB.ER

Unstable time is larger than setting of USER-OTHER-NLD.RNG

PR.TAR

Preset TARE weight

COMP

Input COMPARE data mode

HIGH

Input HIGH limitation data of Comparison

LOW

Input LOW limitation data of Comparison

SPL.Lo

Sample load weight of low point.

SPL.HI

Sample load weight of high point.