63

Table 7.2 Bit Definition of H

1

H

2

H

3

H

4

7.6 Commands and Response

7.6.1 Parameter NCI-SCP01

Set the USER-COM-LYOUT parameter to

SCP01.

W

1

---W

6

Reading data, 1-6 bytes (six digits)

<DP>

Decimal point

U

1

U

2

U

3

U

4

U

5

Measure units, kg, lb, lb:oz, % or pcs; 2-5 bytes

<Add>

Address of scale; 2 bytes (00-99)

<Prompt>

Prompt characters of output content; max. 11bytes

Bit

Byte 1 (H

1

)

Byte 2 (H

2

)

Byte 3 (H

3

)

Byte 4 (H

4

)

0

0 = stable

0 = not under capacity 00 = compare disable

01 = lower limit
10 = ok
11= upper limit

00 = normal weighing
01 = count weighing
10 = percent weighing
11 = other mode

1 = not stable

1 = under capacity

1

0 = not at zero point 0 = not over capacity
1 = at zero point

1 = over capacity

2

0 = RAM ok

0 = ROM ok

0 = gross weight

0 = not in HOLD

1 = RAM error

1 = ROM error

1 = net weight

1 = in HOLD

3

0 = eeprom OK

0 = calibration ok

0 = initial zero ok

0 = battery ok

1 = eeprom error

1 =calibration error

1 = initial zero error

1 = low battery

4

always 1

always 1

always 1

always 1

5

always 1

always 1

always 1

always 1

6

always 0

always 1

always 1

always 0

7

parity

Parity

parity

Parity