6 display software and hardware version, 7 display test, 8 keyboard and buzzer test – Salter Brecknell 6700U Series (USB) User Manual

Page 33

Advertising
background image

67XXU USB/Serial Scale Service Manual

33

The scale will exit HOLD mode according to the setting of USER-HOLD-HLD.TM. If the
time of scale variation being over USER-OTHER-NLD.RG is more than USER-HOLD-
STB.TM,

STB.ER will be shown. Press the [TARE] key to start averaging again or

press the [HOLD] key to exit.

4.6 Display Software and Hardware Version

Display the current version of software and hardware.

1.

Enter the Setup Mode by pressing the [HOLD] + [ON/OFF] keys until

confg

is displayed.

2.

Use the [UNIT] key to select the MISC-VER menu.

3.

Press [TARE] to display the current version.

3a.

Firmware Version display Format is: Vxx.yy, xx is hardware version, yy is
software version.

4.

Press the [ON/OFF] key to return to last menu item.

4.7 Display Test

1.

Press the [HOLD] + [ON/OFF] keys until

confg is displayed.

1.

Use the [UNIT] key to select TEST-DSP.TS menu.

2.

Press [TARE] to enter the test display mode and all segments will light at first.

2a.

In this mode, every press of the [UNIT] key will light the next segment. Every
press of the [HOLD] key will light the next digit.

2b.

Press the [TARE] key to automatically light all segments and all digits.

3.

Press the [ON/OFF] key to return to last menu item.

4.8 Keyboard and Buzzer Test

1.

Press the [HOLD] + [ON/OFF] keys until

confg is displayed.

2.

Use the [UNIT] key to select TEST-KEY.TS menu.

3.

Press [TARE] to enter test keypad mode and

key will be displayed.

4.

Press any key. The value of this key will be displayed and the buzzer will beep.

5.

Press the [ON/OFF] key to return to last menu item.