Filter baud rate/parity 15 – Salter Brecknell 6115 User Manual

Page 18

Advertising
background image

From normal weighing mode press SW-1…

“DIAG” is displayed.

Press the ZERO key until…

“CONF” is displayed.

Press TEST key to access the Configuration menu.

“FILT” is displayed.

Press the TEST key and the present filter param-
eter is displayed. Press the ZERO key to toggle
between choices “FAST” or “SLOW.” When scale is
placed in a area subject to surrounding vibrations
or air (wind) movement, set the filter on “SLOW.”

Press TEST to select the Filter setting and…

“FILT” will again be displayed.

From “FILT” press the ZERO and…

“BAUD” is displayed.

Press the TEST key and the present baud rate and
parity setting are displayed.

Press the ZERO key to scroll through the nine
choices. You can select from 2400, 4800 or 9600
baud with parity settings of even, odd or none.
When the proper selection is displayed, press the
TEST key. The selection is saved and…

“FILT” will again be displayed.

Baud Rate Selections

“24-e”

“24-o”

“24-n”

“48-e”

“48-o”

“48-n”

“96-e”

“96-o”

“96-n”

Filter

Baud Rate/Parity