6. Place chosen span calibration weight on the

scale and wait for the scale to stabilize;
then press the TEST key. The display will flash
dashes and the message…

“DONE” is displayed.

7. Remove calibration weight from the scale.

8. Press the TEST key and the scale performs a

countdown and returns to the normal weigh-
ing mode.

TABLE 1

Alternate Calibration Span Values

Capacity

Alternate Span Calibration Weight

10 lb

2, 5 lb

15 lb

2, 5 lb

25 lb

2, 5 lb

160 oz

32, 80 oz

240 oz

32, 80 oz

400 oz

32, 80 oz

3000 g

1000, 2000g

6000 g

1000, 2000g

10000g

1000g, 2000g

The Units (UNITS) Mode menu lets you change
between pounds and ounces without the use of
calibration weights. The scale must be currently
operating in pounds or ounces to allow a unit of
measure change.

From normal weighing mode, press SW-1…

“DIAG” is displayed.

Press the ZERO key until…

“UNITS” is displayed.

Calibration Span

Requirements

The internal count value of
the calibration test points for
zero and span must provide
a minimum 5% net count
change, or the message
“CAL” “ERROR” “ABORT”
is flashed on the display and
the previous calibration data
is restored. This is done to
insure that a calibration
weight is placed on the
scale for the span value and
to avoid loading of invalid
calibration data into memory.