Missing or extra segments on display
1. Replace the display board.
2. Replace the main board.

Scale will not return to zero, or incorrect weight is
displayed
1. Press the ZERO key.
2. Check for interference of weighing platform.
3. Power down, remove all items from the platter,

and then power up the scale.

4. Recalibrate the scale.
5. Replace the load cell.
6. Replace the main board.

Display shows unrecognized characters
1. Check software PROM for proper insertion.
2. Check display cables for the proper connection.
3. Replace PROM.
4. Replace the main board.

Display shows under “_ _ _ _” dashes

(Indicates that the scale is below zero or under capacity.)

1. Verify that shroud platter is on the scale.
2. Press the ZERO key.
3. Power down, remove any items from the

platter, and then power up the scale.

4. Recalibrate the scale.
5. Replace the load cell.
6. Replace the main board.

Display shows upper”

_ _ _ _

“ dashes

(Indicates the scale is over capacity.)

1. Remove all items from the scale.
2. Press the ZERO key.
3. Power down, and then power up the scale.
4. Recalibrate the scale.
5. Replace the load cell.
6. Replace the main board.