KEYENCE LK-H1W User Manual

Page 77

Advertising
background image

A-5

LK-H1W-M-NOA-E

Appendices

A

* The value in parenthesis ( ) varies depending on the value specified for [Minimum display unit].

Common Settings

Analog scaling output voltage

Illegal entry.
Enter a value in the range of
–10.500 V to +10.500 V.

The analog scaling output
voltage column is left blank. Or
any character other than
numerical value, or a value
exceeding the settable range is
set in the column of analog
scaling output voltage.

Enter a value in the range of
–10.5 V to +10.5 V, for the
analog scaling output voltage.

Overall analog scaling

Implement the scaling setting so
that inclination of the scaling
setting is (9.999) or less per 1 V.

Inclination of the analog scaling
setting exceeds the setting
value.

Set the measured value and
output voltage so that inclination
of the analog scaling setting
becomes smaller than the
displayed value.

Error message

Cause

Remedy

Number of data stored

The settable range is from 1 to
65536.

The Number of data stored field
is left blank or a value of 0 or
larger than 65536 is set in the
field.

Enter a value in the range of 1 to
65536 in the field for Number of
data stored.

Error message

Cause

Remedy

Advertising