12 span calibration to each channel – ADLINK ND-6017 User Manual

Page 57

Advertising
background image

Command Set

49

3.12 Span Calibration to each Channel

@Description

To correct the gain errors of AD converter by using the span calibration.

@Syntax

$(Addr)0(Channel No)<CR>

$

Command leading code (1 character)

(Addr)

Address ID (2 character)

0

Span calibration (1 character)

(Channel No)

Channel for Calibration (1 character) 0~2

@Response

!(Addr)<CR>
or
?(Addr)<CR
>

!

Command is valid.

?

Command is invalid.

(Addr)

Address ID.

@Example

User command:

$0601<CR>

Response:

!06<CR>


To perform the span calibration for analog input module ND-6013
channel 1, address ID is 06H.

Note:

To perform the calibration, a proper input signal should be
connected to the analog input module. Different input range
have different input voltage, detail refer chapter 5
Calibration” .

(for 6013 F/W version
A3.05~A4.60)

Advertising