12 span calibration to each channel – ADLINK ND-6017 User Manual
Page 57
 
Command Set
• 49
3.12 Span Calibration to each Channel
@Description
To correct the gain errors of AD converter by using the span calibration.
@Syntax
$(Addr)0(Channel No)<CR>
$
Command leading code (1 character)
(Addr)
Address ID (2 character)
0
Span calibration (1 character)
(Channel No)
Channel for Calibration (1 character) 0~2
@Response
!(Addr)<CR> 
 or 
?(Addr)<CR
> 
 
!
Command is valid.
?
Command is invalid.
(Addr)
Address ID.
@Example
User command:
$0601<CR>
Response:
!06<CR>
 
To perform the span calibration for analog input module ND-6013 
channel 1, address ID is 06H. 
 
Note:
To perform the calibration, a proper input signal should be 
connected to the analog input module. Different input range 
have different input voltage, detail refer chapter 5 
“Calibration” . 
(for 6013 F/W version 
A3.05~A4.60)