28 cjc offset calibration – ADLINK ND-6017 User Manual
Page 73
                                    Advertising
                                    
                                        
                                    
                                
                                 
Command Set
• 65
3.28 CJC Offset Calibration
@Description
To correct the CJC offset errors use CJC (Cold Junction Compensation) 
offset calibration. 
@Syntax
$(Addr)9(Counts)<CR> 
 
$
Command leading code
(Addr)
Address ID
9
CJC offset calibration.
(Counts)
It is a 4-characters (Hexadecimal) with a sign + or -, 
range is 0000 to FFFF, each count equals 
approximately 0.0153
°C.
Example: +0042 = 4x16 + 2 = 66 
 66 * 0.0153
°C = 1.009°C
@Response
!(Addr)<CR> 
 or 
?(Addr)<CR
> 
 
!
Command is valid.
?
Command is invalid.
(Addr)
Address ID.
@Example
User command:
$089+0042<CR>
Response:
!08<CR>
 
CJC offset calibration at address 08H. The calibrated offset 
temperature is +0042(Hex) = 66, 66 x 0.0153
°C = 1.009°C
(6011/D, 6018)