ADLINK ND-6017 User Manual

Page 95

Advertising
background image

Data Format and Input Range

87

Input Range

Resolution

±15 mV, ±50 mV

1

µV

three decimal places

±100 mV, ±150 mV, ±500 mV

10

µV

two decimal places

±1 V, ±2.50 V, ±5 V

100

µV

four decimal places

±10 V

1mV three

decimal

places

±20 mA

1

µA

three decimal places

Type J and T thermocouple, RTD 0.01

°C

two decimal places

Type K, E, R, S, B, N and C
thermocouple

0.1

°C

one decimal places

Table 4-1 Data format and resolution

Example 1:

Input Range is

±5 V

Input is -1.37 Volts


engineering units: -1.3700<CR>

Example 2:

Input Range is

±10 V

Input is +3.653 Volts


engineering units: +03.653<CR>

Example 3:

Input Range is Type K thermocouple (range 0

°C to 1000°C)

Input is 406.5

°C


engineering units: +0406.5<CR>

Example 4:

Input Range is Type T thermocouple (range -100

°C to 400°C)

Input is -50.5

°C


engineering units:

-

050.50<CR>

Percent of FSR (Full Scale Range)

Data format bit 1 and 0 set to 01 is percent of FSR

This data format including three components

1. sign (+ or - )
2. digits
3. decimal point

Advertising