ADLINK ND-6017 User Manual
Page 95
Data Format and Input Range
• 87
Input Range
Resolution
±15 mV, ±50 mV
1
µV
three decimal places
±100 mV, ±150 mV, ±500 mV
10
µV
two decimal places
±1 V, ±2.50 V, ±5 V
100
µV
four decimal places
±10 V
1mV three
decimal
places
±20 mA
1
µA
three decimal places
Type J and T thermocouple, RTD 0.01
°C
two decimal places
Type K, E, R, S, B, N and C
thermocouple
0.1
°C
one decimal places
Table 4-1 Data format and resolution
Example 1:
•
Input Range is
±5 V
•
Input is -1.37 Volts
engineering units: -1.3700<CR>
Example 2:
•
Input Range is
±10 V
•
Input is +3.653 Volts
engineering units: +03.653<CR>
Example 3:
•
Input Range is Type K thermocouple (range 0
°C to 1000°C)
•
Input is 406.5
°C
engineering units: +0406.5<CR>
Example 4:
•
Input Range is Type T thermocouple (range -100
°C to 400°C)
•
Input is -50.5
°C
engineering units:
-
050.50<CR>
Percent of FSR (Full Scale Range)
•
Data format bit 1 and 0 set to 01 is percent of FSR
•
This data format including three components
1. sign (+ or - )
2. digits
3. decimal point