ADLINK ND-6017 User Manual
Page 97
 
Data Format and Input Range
• 89
 
Two’s complement hexadecimal: CD27<CR> 
((-2/5) x 32768) = -13107.2 = CD27H 
 
Example 3:
•
Input Range is
±10 V
•
Input is +4 Volts
 
Two’s complement hexadecimal: 3333<CR> 
 ((4/10) x 32768) = 13107.2 = 3333H 
 
Example 4:
•
Input Range is Type K thermocouple (range 0
°C to 1000°C)
•
Input is 406.5
°C
 
Two’s complement hexadecimal: 3408<CR> 
 ((406.5/1000) x 32768 ) = 13320.2 = 3408H 
Ohm
•
Data format bit 1 and 0 set to 11 is ohm presentation.
•
This data format including three components.
 1. sign (+) 
 2. digits 
 3. decimal point 
• 
Data is sign (+) followed with 5-digits and a decimal point
•
It does not exceed 7-characters
•
Maximum resolution is 0.01ohm, the decimal point is fixed
 
Example 1: 
•
Input Range is Pt-100, -100
°C to +100°C, α=0.00385
•
Input is 120.23 ohm
 
ohm: +120.23<CR>