Flowcharts

Product Reference Guide

4-11

Figure 4-6. Problem Isolation: Scale

SCALE START

View the Remote
Display and/or the
Yellow scale status LED
for the scale error code.
Reference the topic,
Scale Error Reporting in
Section 4 for more
information.

Problem is too much
motion at power-up.
Correct excess vibration
in counter if possible.

Increase the scale
motion filter setting by
1. Reference Section 6
for more information.

NO

YES

E__1

shown on

Remote Display/

Yellow LED

Flash once

?

Problem is calibration
lost. Recalibrate the scale
using the procedures in
Section 5.

NO

YES

E__2

shown on

Remote Display/

Yellow LED

Flash twice

?

- Check for debris under platter.
- Verify that the weigh platter
moves freely.
- Remove item(s) from scale.
- Press Scale Zero Button
- If the scale still fails to zero,
recalibrate the scale.

NO

YES

-0-

shown on

RemoteDisplay/

Yellow

LED off

?

NO

NO

NO

YES

Is

problem

fixed

?

NO

Is

problem

fixed

?

NO

YES

YES

Is the

motion filter

at its maximum

setting

?

DONE

YES

Will the

scale

calibrate

?

NO

YES

E__3

shown on

RemoteDisplay/

Yellow LED Flash

3 times

?

E__4 shown on Remote
Display/Yellow LED Flash
4 times. Problem is Scale
Load Cell failure.

YES

Is

problem

fixed

?

DONE

DONE

DONE

Call Tech Support

Call Tech Support

If this is a new installation,
make sure any foam packing
pieces have been removed
from the scale cavity before
proceeding. Lift off the platter
to verify, then set the platter
back into position.

For this error, the 7-segment
display may also show a 'C'.

Problem is scale
communication lost.
Verify cable connections
to/from the scale.