Warm-up time, Warm-up time -8 – Datalogic Scanning 9500 User Manual

Page 20

Advertising
background image

1-8

Magellan

®

9500

Warm-Up Time

There are two pertinent warm-up times that apply to the scanner or scan-
ner/scale:

Thermal Equilibrium

When the unit is moved from a cooler temperature (such as a storage area)
to a warmer environment (such as a checkstand location), 60 minutes
must be allowed to acclimate the unit to ambient conditions prior to cali-
bration or operation.

Power-up

Once installed and powered up, a warm-up time of 15 minutes must be
allowed before calibrating or performing weighing operations.

User Configurable Warm-up

The user may configure the unit for a pre-programmed warm-up time
that is activated every time the scanner is powered up. During this time,
the scale is viewed by the POS terminal as off-line.

NOTE

The two warm-up periods can be performed concurrently, thereby reducing
the total required warm-up time to 60 minutes.

NOTE

Contact technical support to learn more about this advanced program-
mable feature.