VEGASON 51 V … 53 V

13

Types and versions

-

Configuration of measuring systems

2

2

4

4

VEGAMET

515V

2

Processings see also product information
"Signal conditioning instruments series 500"

Current outputs
Voltage outputs
Relays
Digital connection
Fault signals

Screened line in case of electro-
magnetic interferences

VEGACONNECT 2

Signal conditioning instru-
ment VEGAMET 515 V in
housing type 505

VEGADIS 50

1 … 2 sensors on signal conditioning instrument VEGAMET 515 V

Two-wire technology, supply from signal conditioning instrument. Output
signals and voltage supply via one two-wire line.

Digital output signal, two sensors on one line.

Measured value indication in sensor and in signal conditioning instrument.

Optionally external indicating instrument.
(can be mounted up to 25 m separated from the sensor in Ex-area).

Adjustment with PC, signal conditioning instrument or adjustment module
(can be plugged in sensor or in external indicating instrument).

Max. resistance of the signal line 15

per wire or 1000 m

cable length

VEGADIS 50

1) Sensor lines should be looped in screened

cables. It is suitable to earth the cable
screens on both ends. However it must be
noted that no earth compensation currents
flow via the screens.
Earth compensation currents can be avoided
with earthing on both ends by connecting the
cable screen on one earth end (e.g. in the
switching cabinet) via a capacitor (e.g. 1

µ

F,

100 V) to the earth potential.