3 technical data, 1 data – VEGA VEGASON 51V…53V User Manual

Page 17

Advertising
background image

VEGASON 51 V … 53 V

17

Technical data

3 Technical data

3.1 Data

Type

51…

52…

53…

Power supply

Supply voltage

from signal conditioning instrument VEGAMET or
processing system VEGALOG 571 (max. 36 V DC)
fuse 0,5 A

(slow-blow)

Current consumption

max 22,5 mA

Power consumption

max. 80 mW

0,45 VA

Meas. range (relating to the transducer diaphragm)

Liquid

0,25 … 4 m

0,4 … 7 m

0,6 … 15 m

Solid

0,30 … 2 m

0,5 … 3,5 m

0,75 … 7 m

Output signal

Digital meas. signal (VBUS)

Adjustment

- PC with adjustment software software VEGA Visual Operating
- adjustment module MINICOM

Accuracy (type, values under reference conditions)

1)

Linearity error after adjustment

< 0,1 % (relating to max. meas. range)

Average temperature error of sensor
electronics

< 0,03 %/10 K of meas. range

Resolution

1 mm

Characteristics

Min. span
(between empty and full adjustment)

> 10 mm (recommended > 50 mm)

Ultrasonic frequency

70 kHz

55 kHz

38 kHz

Meas. intervals

1,0 s

1,0 s

0,6 s

Beam angle

5,5

°

5,5

°

3

°

(at -3 dB emitted power)

1) Reference conditions acc. to IEC 770, e.g.:

Temperature:

18 ... 30

°

C

etc.