VEGA VEGASON 51V…53V User Manual

Page 14

Advertising
background image

14

VEGASON 51 V … 53 V

Types and versions

-

Configuration of measuring systems

2

2

4

4

VEGAMET

515V

VEGATRENN

547

2

Ex-area

Not Ex-area

Processings see also product information
"Signal conditioning instruments series 500"

Screened line in case of electro-
magnetic interferences

VEGACONNECT 2

Signal conditioning instrument VEGAMET
515 V with Ex-separator VEGATRENN
548 V Ex in housing type 506

Current outputs
Voltage outputs
Relays
Digital wiring
Fault signals

VEGADIS 50

VEGADIS 50

1 … 2 sensors in Ex-area via separator VEGATRENN 548 V Ex on signal condi-
tioning instrument VEGAMET 515 V

Two-wire technology, supply from the signal conditioning instrument. Output
signals and voltage supply via one two-wire line (optionally with
VEGASON 53 four-wire technology)

VEGASON 51 … 52:

CENELEC EEx ia IIC T6

VEGASON 53:

StEx Zone 10

Digital output signal, two sensors on one line.

Measured value indication in the sensor and in the signal conditioning in
strument.

Optionally external indicating instrument
(can be mounted up to 25 m separated from the sensor in Ex-area).

Adjustment with PC, the signal conditioning instrument or the adjustment
module (can be plugged into the sensor or the external indicating instru
ment).

Max. resistance of the signal line 15

per wire or 1000 m cable length note

the permissible values for capacitance and inductance at the connection
cable (see approval certificate of the separators)

Advertising