VEGA VEGASON 51V…53V User Manual
Page 58
58
VEGASON 51 V … 53 V
Backup
• Click to
"Services/Backup/Sensors"
In the menu window "
Backup" the sensor with
the serial number is displayed. You can save
the sensor individually or in groups with all
adjustments in a directory of your choice on
your PC. In addition you can add a short note
to each backup.
You also save the adjustments of the signal
conditioning instrument VEGAMET:
Saved sensor data can be transferred later
to other sensors. If you have e.g. a system
with several of the same storage vessels and
identical sensors, it is sufficient to configure
one sensor and then transfer to the other
sensors.
• Choose the menu "Services/Restore con-
figuration/Sensors"
In this menu window you choose in the yellow
window cutout a sensor (or a signal condi-
tioning instrument) which should be
overwritten by the adjustments of another
sensor.
Choose in the field "
Selection of backups" the
serial number of the sensor of which you
want to transfer the adjustments. With "
Re-
store to" you transfer these sensor adjust-
ments to the sensor you have chosen in the
yellow window cutout.
Set-up
-
Adjustment with PC on VEGAMET