Configuration of the input, ) operating range, ) measuring conditions – VEGA VEGASON 51V…53V User Manual

Page 61: ) adjustment

Advertising
background image

VEGASON 51 V … 53 V

61

Adjustment without medium

Key adjustment

Display indication

The distance-indication
flashes slowly

Now you can enter with the "

+" and "–" key

the distance, which your sensor has to the
product at 0 % filling (example: 5,850 m)

Adjust
ment

Param.
adjust
ment

Sensor

m(d)

4.700

w.out
medium

Adjust
ment
in

m(d)

0.0 %

at

m (d)

XX.XXX

(Min. adjustment)

OK

OK

+

OK

OK

Set-up

-

Adjustment with VEGAMET or MINICOM

Following short explanations to the set-up
steps 1 … 9.

1. Configuration of the input

First of all you have to inform the signal condi-
tioning instrument (only VEGAMET 515 V) to
which input (input 1 or input 2) the sensor is
connected.

2.) Operating range

Without special adjustment, the operating
range corresponds to the meas. range. The
meas. range has already been adjusted with
the min./max. adjustment. Generally it is use-
ful to choose the operating range slightly
bigger (approx. 5 %) than the meas. range.
Example:
Min./max. adjustment: 0,500 … 3,400 m;
adjust operating range to approx.
0,400 … 3,500 m.

3.) Measuring conditions

see menu plan figure 3 on page 70.

4.) Adjustment

Under the menu point "

Adjustment" you inform

the sensor in which meas. range it should
operate.

The adjustment can be carried out with and
without medium. Generally you will carry out
the adjustment without medium as you can
adjust without filling cycle.

Max.

Min.

100 %

(0,500 m product distance)
correspond to1200 litres

Meas. range
0,5…3,4 m

0 %
(3,400 m product distance)
correspond to 45 litres

Advertising