100 mm – VEGA VEGADIF 55 4 … 20 mA_HART User Manual

Page 22

Advertising
background image

l

R

emember that due to the hydrostatic pressure of the liquid

columns in the capillaries, a zero point shift can occur. The zero
point shift can be corrected, see also "Position adjustment".

l

K

eep the application limits of the isolating diaphragm filling oil in

mind according to the Product Information manual VEGADIF 55,
"

Planning instructions isolating diaphragms"

Capillaries
T

o achieve more precise measuring results and avoid instrument

malfunctions, the capillaries must be mounted as follows:

Note:

l

V

ibration-free (to avoid additional pressure fluctuations)

l

N

ot close to heating or cooling lines

l

I

nsulate them in cooler or warmer environment

l

L

ay wire with a bend radius ≥ 100 mm

l

W

ith isolating diaphragms on both sides, the ambient temperature

and length of both capillaries should be the same

l

A

lways two identical isolating diaphragms (e.g. diameter, material

etc.) should be used for the minus and plus side (standard
shipment)

+ –

+

≥ 100 mm

Fig. 18: Mounting VEGADIF 55 with isolating diaphragms and capillaries with
vacuum application: place the pressure transmitter below the lower isolating
diaphragm