DIP switch

6

.7 Copy configuration data

Information:
T

o copy configuration data from the data memory module in an

instrument, the adjustment must be unlocked.

Copy configuration data from one instrument into a data memory
module:

Information:
T

he adjustment must be unlocked. See also chapter "Lock/Unlock

adjustment".

1

S

eparate instrument from voltage supply

2

P

lug data memory module into the electronics module

3

C

onnect instrument again to voltage supply

4

P

ush "E" and "-" keys (at least 3 seconds) until the LED on the

electronics module lights

5

W

ait approx. 20 seconds. Configuration data are loaded from the

instrument to the data memory module. The instrument does not
carry out a restart.

6

S

eparate instrument again from voltage supply

7

R

emove memory module

8

C

onnect instrument again to voltage supply

Local indication not
connected

VEGADIF

55 • 4 … 20

mA/HART

45

6 O

peration

31731

-EN

-081119