VEGA VEGADIF 55 4 … 20 mA_HART User Manual
Page 51
Reset code
Description
7864
Total reset
T
his reset resets the following parameters:
l Function group POSITION ADJUSTMENT
l Function group BASIC SETUP
l Function group EXT. SETUP
l Function group LINEARIZATION (an existing linearization chart may be deleted)
l Function group: TOTALIZER SETUP
l Group OUTPUT
l Function group HART DATA
l Function group MESSAGES
l All messages to be configured (type "Error") are set to "Warning", see also Supplementary
instruction manual "Service instructions VEGADIF 55", "Messages" and chapter "Fault
rectification"
l Function group USER LIMITS
l Function group SYSTEM 2
l If a simulation is running, this will be terminated
l The instrument carries out a restart
8888
Data memory module reset
M
easured value and event memory will be deleted. The data memory module must be plugged to
the electronics module during the reset.
E
xample LEVEL MODE = linear and LIN. MEASURAND = Height
l
HEIGHT UNIT =
m
l
CALIBRATION MODE =
wet
l
EMPTY CALIB. = 0
l
FULL CALIB. = S
ensor final value converted in H²O, e.g. with a
500
mbar sensor: 5.99 mH²O
VEGADIF
55 • 4 … 20
mA/HART
51
6 O
peration
31731
-EN
-081119