VEGA VEGADIF 55 4 … 20 mA_HART User Manual

Page 37

Advertising
background image

l

"+" key and "-" key

-

Q

uit editing mode without saving the modified value

-

I

n the menu you are within a function group: when pushing the

keys for the first time simultaneously, you jump one parameter

back within the function group. With every further simulta-
neous pushing you jump one level in the menu to the top

-

Y

ou are in the menu on a selection level: With every

simultaneous pushing you jump one level in the menu to the
top

l

DIP

switch 1

-

T

o lock/unlock measured value-relevant parameters. Default

setting: off (unlocked)

l

DIP

switch 2

-

D

amping on/off, default setting: on (damping on)

6

.4 Local adjustment - local indication not

connected

Information:
F

or adjustment of the instrument with a data memory module see

"

Data memory module".

I

f no local indication is connected, the following functions will be

possible via the three keys on the electronics module or outside on the
instrument:

l

P

osition adjustment (zero point correction)

l

S

et measurement begin and measurement end

l

I

nstrument reset, see chart in chapter "Operating elements"

Information:
T

he adjustment must be unlocked, see chapter "Lock/Unlock adjust-

ment".

B

y default, the instrument is set to mode Pressure. The mode can be

changed via the parameter "Measuring Mode", see chapter "Select
language and measuring mode
".

T

he existing pressure must be within the nominal pressure limits of the

sensor, see specifications on the "Type label".

Carry out position adjustment:

1

P

ressure on the instrument

2

P

ush "E" key at least for 3 s

3

D

oes the LED on the electronics module light briefly?

Y

es: existing pressure was accepted as position adjustment

N

o: Existing pressure for position adjustment was not accepted,

note the adjustment limits

Set LRV:

Measuring Mode - Pres-
sure

VEGADIF

55 • 4 … 20

mA/HART

37

6 O

peration

31731

-EN

-081119