Fig. 66: Closed vessel
I
VEGADIF 55
II
Three-fold valve block
III
Precipitator
1
, 5 Drain valves
2
, 4 Inlet valves
6
, 7 Vent valves on VEGADIF 55
A, B Block valves
Proceed as follows to prepare for the actual adjustment:
1
F
ill the vessel to just above the lower tap
2
F
ill measuring system with medium
C
lose valve 3: Separate plus and minus side
O
pen valve A and B: Open block valves
3
V
ent plus side (probably empty minus side)
O
pen valve 2 and 4: Discharge medium on the plus side
B
riefly open valve 6 and 7, then close again: Fill the plus side
completely with the medium and remove air.
4
S
et measurement loop to operation
N
ow:
V
alve 3, 6 and 7 are closed
V
alve 2, 4 A and B are open
5
C
arry out adjustment, see end of chapter "Level measurement"