7 switch on phase – VEGA VEGADIF 55 4 … 20 mA_HART User Manual

Page 31

Advertising
background image

l

I

s the instrument connected according to chapter "Connecting to

power supply"?

l

D

id you tighten all screws?

l

D

id you screw up the housing covers?

T

he green LED on the electronics module lights for a few seconds or

the connected local indication lights if the voltage is switched on on the
instrument.

5

.7 Switch on phase

A

fter VEGADIF 55 is connected to voltage supply or after voltage

recurrence, the instrument carries out a self-check. The following
steps are carried out:

l

I

nternal check of the electronics

l

F

or approx. 3 seconds, the output signal takes on the value 12 mA

T

hen the actual measured value is displayed and the corresponding

current is transmitted to the cable.

1)

1)

T

he values correspond to the actual measured value as well as to the

settings already carried out, e.g. default setting.

Switch on phase

VEGADIF

55 • 4 … 20

mA/HART

31

5 C

onnecting to power supply

31731

-EN

-081119

Advertising