.2 Connection procedure

P

roceed as follows:

1

U

nscrew the housing cover

2

L

oosen compression nut of the cable entry

3

R

emove approx. 10 cm (4 in) of the cable mantle, strip approx.

1

cm (0.4 in) insulation from the ends of the individual wires

4

I

nsert the cable through the cable gland into the sensor

5

L

oosen the screws on the terminals

6

I

nsert the wire ends into the open terminals according to the wiring

plan

7

T

ighten the screws of the terminals

8

C

heck the hold of the wires in the terminals by lightly pulling on

them

9

C

onnect the screen to the internal ground terminal, connect the

outer ground terminal with potential equalisation

10 T

ighten the compression nut of the cable entry. The seal ring must

completely encircle the cable

11 S

crew the housing cover on

T

he electrical connection is finished.

5

.3 Wiring plan

Test

Test

4... 20mA

5

4

3

2

1

Fig. 24: Housing overview, view to the connection compartment
1

Housing

2

Bridge for 4 … 20 mA test signal

3

Internal ground terminal

4

External ground terminal

5

4

… 20 mA test signal between plus and test terminal

Housing overview