46

Operatingandfunctions

When the device generates a warning message, the device status LED is orange.

The generation of a warning message can also be indicated by the switching of the relay output.

OUtPUt

DO2

wARNiNG

DO3

iNV YEs

iNV NO

0......9

Choose the operating, inverted or not inverted, of the
relay output.

DEL. 2= 00

Enter the time delay before switching value (value
between 0 and 99 s).

Confirm.

Fig. 42:

Configuration of the DO2 or DO3 relay output to indicate the generation of a warning message

If you do not want to adjust another parameter, go to the "END" parameter of the Parameters menu and press

ENTER

to save the settings or not and go back to the Process level.

8.6.12. configuring the DI1 digital input

The DI1 digital input makes it possible to remotely trigger one out of four device functions.

iNPUt

Di1

0......9

DisABLE

CALiB. 0

hOLD

sEt FLOw

REs. tOt.

REtURN

Choosing not to use the digital input.

Configuring the digital input to freeze the flow rate
measurements. See “Fig. 45”.

Configuring the digital input to control the status of
the outputs and the display depending on a preset
flow rate. See “Fig. 46”.

Configuring the digital input to trigger the reset of the
daily totalizer. See “Fig. 47”.

Configuring the digital input to trigger the calibration
of the flow zero point. The calibration can also be
done in the Test menu. See “Fig. 44”.

0......9

Fig. 43:

Diagram of the "InPUT" parameter of the Parameters menu