Configuring the electric network frequency, Parameterizing the cut-off flow rate – Burkert Type 8045 User Manual

Page 55

Advertising
background image

51

Operatingandfunctions

8.6.15. configuring the electric network frequency

This parameter makes it possible to configure the electric network frequency so that the device can filter the inter-
fering signals of the power supply.

Adjust this parameter even if the device operates with a DC voltage.

FREqUENC.

F = 50hz

CUt-OFF

0......9

Choose the electric network frequency.

Confirm

F = 60hz

Fig. 51:

Diagram of the "FREQUEnC." parameter of the Parameters menu

If you do not want to adjust another parameter, go to the "END" parameter of the Parameters menu and press

ENTER

to save the settings or not and go back to the Process level.

8.6.16. parameterizing the cut-off flow rate

This parameter makes it possible to set the flow rate value to 0 if the measured value is less than the set cut-off
value:

• the display then shows a flow rate = 0 (a dot is displayed after the flow rate units).

• the outputs and the totalizers react as if the actual flow rate was = 0.

CUt-OFF

CO=0.000

BACKLit

Enter the measured flow rate value below which the device considers that the flow rate
is = 0.

To disable the function, set CO=0.

Confirm

Fig. 52:

Diagram of the "CUT-OFF" parameter of the Parameters menu

If you do not want to adjust another parameter, go to the "END" parameter of the Parameters menu and press

ENTER

to save the settings or not and go back to the Process level.

English

Type 8045