Calibrating the flow zero point – Burkert Type 8045 User Manual

Page 59

Advertising
background image

55

Operatingandfunctions

8.7.2.

calibrating the flow zero point

Adjust this parameter:

• before carrying out a teach-in procedure of the K factor.

• after maintenance work.

• if the measured flow rate is not zero whereas the fluid circulation has been stopped.

Make sure there are no bubbles in the pipe.

Before calibrating the flow zero point:

• at first commissioning the device, immerse the sensor in fluid for 24 hours before calibration.

• after maintenance work, immerse the sensor in the fluid for 1 hour before calibration.

During the calibration:

• the device status LED flashes.

• the outputs are frozen to the last measured flow rate value.

• the device cannot be configured.

The flow zero point calibration can be:

• either remotely triggered via the digital input. See chap. “8.6.12” and Fig. 56.

• either carried out using this parameter in the Test menu; See Fig. 57.

0.000 L/s

Charge the pipe.

Stop the fluid circulation.

Make sure the fluid is still.

Make sure the device status LED is green.

Make sure the device is in the Process level and not in the Configuration level.

Activate the digital input configured to trigger the flow zeo point calibration.

mEAsURE /

The device calibrates the flow zero point during approximately
30 seconds.

• if the device status LED remains green, the calibration of the flow zero point has

succeeded.

• if the device status LED is orange, the calibration of the flow zero point failed: the

message "CAL. FAIL" is displayed in the warning message list. Restart the calibration.

Process level

Fig. 56:

Flow zero point calibration via digital input

English

Type 8045

Advertising