Checking the outputs behaviour – Burkert Type 8045 User Manual

Page 60

Advertising
background image

56

Operatingandfunctions

CALiB. 0

CALiB N

FLOw

Charge the pipe.

Stop the fluid circulation.

Make sure the fluid is still.

mEAsURE /

The device calibrates the flow zero point.

CALiB Y

Choose "CALIB Y".

0......9

The calibration has
succeeded.

The device has detected a non nil

flow rate in the pipe.

The device status LED becomes

orange.

CAL. FAiL

ENTER

Fig. 57:

Flow zero point calibration using the "CALIB 0" parameter of the Test menu

If you do not want to adjust another parameter, go to the "END" parameter of the Test menu and press

ENTER

to save the settings or not and go back to the Process level.

8.7.3.

checking the outputs behaviour

This parameter makes it possible to check that the outputs are behaving as expected by the configuration made.

• The totalizers are incremented depending on the measured value of the flow rate and not the simulated

value.

• The device status LED flashes during the running check of the output behaviour.

English

Type 8045