MAP Technology

Using Luminex 100 IS 2.3 Software

PN 89-00002-00-072 Rev. C

5 - 65

Figure 5-33. Edit Patient List Dialog Box

Assign Sample

Dilution Factors

Indicate the sample dilution factor so the sample analysis in
quantitative tests is accurate. The system multiplies the result by the
dilution factor for reporting.

Do not use dilution factors for qualitative testing.

You define the dilution factors on a sample-by-sample, or patient-by-
patient basis. The list displays each sample’s accession number and
dilution factor in relation to its well position.

‹

To assign sample dilution factors:

1. On the File menu, click New Batch. The Open Template dialog

box opens.

2. Select the desired template and click Select. The Luminex

Batch Setup dialog box opens. See Figure 5-34.

3. Select the item that uses the dilution factor that you are setting.

You cannot change any item that is grayed out or locked into the
command list.