Troubleshooting

PN 89-00002-00-072 Rev. C

7 - 13

Error States

The following error states display on the Status Bar on the Run Batch
tab when an error occurs during system operation.

System Error

Messages

There are two types of error messages: system error messages, and
sample error messages.

System error messages are displayed in three places:

• the Message Log on the Diagnostics Tab

• the message log located in the Windows message log folder

• the Errors tab

You are using incompatible
sheath fluid.

Use only Luminex sheath fluid
in the Luminex 100 analyzer.
Other fluids may damage your
analyzer.

Error Message

Possible problem

Solution

Disconnected

The software hasn’t made initial
connection.

On the toolbar, click Connect;
wait for the status to change.

The communication cables are
not connected.

Remove and reconnect the cable
connections.

Close the software application
and reopen it.

Close the software application,
then turn off the Luminex
system. Power on the system
and check the status.

Refill Sheath

Sheath fluid is low.

Refill the sheath container, then
click Resume.

Running: Sheath Empty

Sheath fluid is empty.

Replace or fill the sheath fluid
container with sheath fluid. Run
two Prime commands.