Drain the analyzer -94 – Luminex 100 IS Version 2.3 User Manual

Page 136

Advertising
background image

Luminex 100 IS User Manual Version 2.3

x

MAP Technology

5 - 94

PN 89-00002-00-072 Rev. C

Drain the Analyzer

When draining, you do not need to supply solution. Draining takes
approximately two minutes and should be followed by an alcohol
flush with 70% isopropanol or 70% ethanol.

Any fluid that drains from the system drains to the Luminex XYP
reservoir as the default. However, you can set the system to drain to
any unused well on the microtiter plate. The drain function normally
expels 125 µL of fluid. Ensure that the location you select to expel
fluid has the reserve capacity to hold the volume expelled.

Sanitize

To decontaminate sample lines and cuvette
after biohazard contact using 10% to 20%
household bleach

– daily if working with biohazards

– monthly if not working with biohazards

Wash

As needed using distilled water

Four times after system calibration

Twice after sanitize

Drain

For troubleshooting and preventative
maintenance purposes only:

– drain the cuvette and refill in preparation for

running. Draining the system removes
debris from the bottom of the cuvette.

– when draining, you do not supply solution.

Draining takes approximately two minutes
and should be followed by an alcohol flush
using 70% isopropanol or 70% ethanol.

Soak

Daily, at the end of the day for shutdown

To prevent salt crystals from forming in the
probe due to exposure to air. Soaking the
probe replaces sheath fluid in the probe with
water. The system uses at least 250 µL of
distilled water.

Self
Diagnostics

To verify system operation.

Table 5-11. Maintenance Operations: