Run system xmap controls -49 – Luminex 100 IS Version 2.3 User Manual

Page 91

Advertising
background image

x

MAP Technology

Using Luminex 100 IS 2.3 Software

PN 89-00002-00-072 Rev. C

5 - 49

You must run system controls following calibration. Continue with
the following “Run System xMAP Controls” section.

Run System xMAP

Controls

Run System xMAP controls to verify calibration. Dispense the
Control microspheres into the microtiter plate at the same time that
you dispense the Calibration microspheres. See “Run System xMAP
Calibrators” on page 5-47
.

‹

To verify system calibration with controls:

1. In the Maintenance tab, click New CON Targ. The Update

CON Target Information dialog box opens. See Figure 5-26.

2. Enter the CON1 lot number, expiration date, and the values

listed from the COQ included with your system controls into the
CON1 entry boxes. If you are using a previously entered lot,
select it from the drop-down menu in the Current CON 1 group
box or click Import to import the information. See page 5-50 for
more details on importing a lot.

3. Repeat step 2 for the CON2 lot information.

Figure 5-26. Update CON Targets Dialog Box