Luminex 100 IS Version 2.3 User Manual

Page 140

Advertising
background image

Luminex 100 IS User Manual Version 2.3

x

MAP Technology

5 - 98

PN 89-00002-00-072 Rev. C

Perform a Soak

Command

Soak the sample probe to prevent the sheath fluid crystals from
forming in the sample probe.

‹

To perform a soak command:

1. On the Maintenance tab, click Eject/Retract.

2. Select Reservoir from the dropdown menu next to the Soak

button, then click Soak. A confirmation dialog box opens telling
you to place solution in the reservoir.

3. Put distilled water in the reservoir.

4. Click OK. The plate holder retracts, and the system performs the

Soak command.

Exit Luminex IS

2.3 Software

When you exit the system a confirmation dialog box prompts you to
verify that you really want to exit the system.

‹

To exit the system:

On the File menu, click Exit, then click Yes.