Xmap technology reagents, Required laboratory reagents, Luminex 100 is 2.3 software – Luminex 100 IS Version 2.3 User Manual

Page 29: Luminex 100 is performance specification, Speed, Speed -3

Advertising
background image

x

MAP Technology

The System

PN 89-00002-00-072 Rev. C

3 - 3

xMAP Technology

Reagents

Classification calibration microspheres (CAL1)

Reporter calibration microspheres (CAL2)

Classification control microspheres (CON1)

Reporter control microspheres (CON2)

Sheath fluid

Required

Laboratory

Reagents

Household bleach

70% isopropanol or 70% ethanol

Mild detergent

Luminex 100 IS 2.3

Software

Luminex 100 IS 2.3 software provides complete control of the
system and performs data analysis. Your Luminex 100 IS 2.3 system
is preloaded with the Luminex software. However, we supply a
software CD should you need to reinstall the software.

This software requires a dedicated system. Unauthorized additional
software is prohibited and may result in improper operation of the
system.

Luminex 100 IS

Performance

Specification

Speed

USB communications link for fast data transfer

Automatic transfer of assay templates and new reagent
information into the system via a large capacity read/write CD

Installation: < 4 hours

System calibration: < 10 minutes

System controls: < 10 minutes

Barcode reader entry of sample IDs

Automatic post-analysis

Analyze one 96-well plate/hour depending on manufacturer’s kit

Up to 100 xMAP microsphere sets per sample

Sheath flowrate: 90 µL/sec ± 5 µL

Sample injection rate into detector area: 1 µL/sec ± 0.05 µL

System warmup: 30 minutes. Systems that remain inactive for at
least four hours will require a warm-up to restart the lasers. After